Effectus:
Modus ingressus: praesidium fluctuans et defibrillationis
Frequentiae responsio: 0.05Hz ~ 150Hz (-3dB ~ +0.4dB)
CMRR: >60dB, >100dB (filtrum adde)
Constans temporis: ≥ 3.2s
Fluxus fluxus aegroti: <10µA
Tensio calibrationis: 1mV
Sensibilitate: 5, 10, 20 mm/mV ±5%, sensibilitate ordinaria: 10 mm/mV ±2%
Gradus sonitus: ≤15µVp-p
Cursus circuitus ingressus: ≤50nA
Impedentia ingressus: ≥50MΩ
Accuratio exempli: 12-bit
Classificatio securitatis: classis I, typus CF et pars applicata defibrillationi immune
Modus inscriptionis: systema impressionis thermalis
Parametri mensurae: FC, intervallum PR, Duratio P, Duratio QRS, Duratio T, intervallum QT, Q-Tc, Axis P, Axis QRS, Axis T, R(V5), S(V1), R(V5) + S(V1)
Magnitudo chartae: 50mm (Latitudo) × 20m (Longitudo)
Frequentia exemplificationis: 1000Hz
Tensio polarizationis: ±500mV
Filtrum interferentiae EMG: 25Hz/35Hz (-3dB)
Filtrum frequentiae potentiae: AC 50Hz/60Hz (-20dB)
Celeritas chartae:
Automaticum: 6.25, 12.5, 25, 50mm/s, ±5%
Manuale: 6.25, 12.5, 25, 50mm/s, ±5%
Fons potentiae: Adaptator DC 12V, accumulator lithii rechargeabilis 7.4V/2000mAh
Gradus impermeabilitatis: IPX0
Modus operandi: labor continuus